Measurement Accuracy |
Voltage DC |
Range/ Resolution |
1 Year |
2 Years |
100.000 mV |
0.02% + 0.005 mV |
0.03% + 0.005 mV |
3.00000 V |
0.02% + 0.00005 V |
0.03% + 0.00005 V |
30.0000 V |
0.02% + 0.0005 V |
0.03% + 0.0005 V |
300.00 V |
0.05% + 0.05 V |
0.07% + 0.05 V |
Voltage AC |
3.000 V (40 Hz to 500 Hz) / 0.001 V |
0.5% + 0.002 V |
1.0% + 0.004 V |
30.00 V (40 Hz to 500 Hz) / 0.01 V |
0.5% + 0.02 V |
1.0% + 0.04 V |
300.0 V (40 Hz to 500 Hz) / 0.1 V |
0.5% + 0.2 V |
1.0% + 0.2 V |
Current DC |
30.000 mA |
0.01% + 5 uA |
0.015% + 7 uA |
110.00 mA |
0.01% + 20 uA |
0.015% + 30 uA |
Resistance |
10.000 Ω |
0.05% + 50 mΩ |
0.07% + 70 mΩ |
100.00 Ω |
0.05% + 50 mΩ |
0.07% + 70 mΩ |
1.0000 kΩ |
0.05% + 500 mΩ |
0.07% + 0.5 Ω |
10.000 kΩ |
0.1% + 10 Ω |
0.15% + 15 Ω |
Frequency |
1.00 to 110.00 Hz / 0.01 Hz |
|
0.05 Hz |
110.1 to 1100.0 Hz / 0.1 Hz |
|
0.5 Hz |
1.101 to 11.000 kHz / 0.001 kHz |
|
0.005 kHz |
11.01 to 50.00 kHz / 0.01 kHz |
|
0.05 kHz |
Source Accuracy |
|
|
1 Year |
2 Years |
Voltage DC |
100.000 mV |
0.01% + 0.005 mV |
0.015% + 0.005 mV |
1.00000 V |
0.01% + 0.00005 V |
0.015% + 0.0005 V |
15.0000 V |
0.01% + 0.0005 V |
0.015% + 0.0005 V |
Current DC |
22.000 mA (source) |
0.01% + 0.003 mA |
0.02% + 0.003 mA |
Current sink (simulate) |
0.02% + 0.007 mA |
0.04% + 0.007 mA |
Resistance |
10.000 Ω |
0.01% + 10 mΩ |
0.015% + 15 mΩ |
100.00 Ω |
0.01% + 20 mΩ |
0.015% + 30 mΩ |
1.0000 kΩ |
0.02% + 0.2 Ω |
0.03% + 0.3 Ω |
10.000 kΩ |
0.02% + 3 Ω |
0.03% + 5 Ω |
Frequency |
0.1 to 10.99 Hz |
|
0.01 Hz |
0.01 to 10.99 Hz |
|
0.01 Hz |
11.00 to 109.99 Hz |
|
0.1 Hz |
110.0 to 1099.9 Hz |
|
0.1 Hz |
1.100 to 21.999 kHz |
|
0.002 kHz |
22.000 to 50.000 kHz |
|
0.005 kHz |
Technical Data |
Data log functions |
Measure functions |
Voltage, current, resistance, frequency, temperature, pressure |
Reading rate |
1, 2, 5, 10, 20, 30, or 60 readings/minute |
Maximum record length |
8000 readings (7980 for 30 or 60 readings/minute) |
Ramp functions |
Source functions |
Voltage, current, resistance, frequency, temperature |
Rate |
4 steps/second |
Trip detect |
Continuity or voltage (continuity detection not available when sourcing current) |
Loop power function |
Voltage |
Selectable, 26 V |
Accuracy |
10%, 18 V minimum at 22 mA |
Maximum current |
25 mA, short circuit protected |
Maximum input voltage |
50 V DC |
Step functions |
Source functions |
Voltage, current, resistance, frequency, temperature |
Manual step |
Selectable step, change with arrow buttons |
Autostep |
Fully programmable for function, start delay, stepvalue, time per step, repeat |
Environmental Specifications |
Operating temperature |
-10°C to +50°C |
Storage temperature |
-20°C to +60°C |
Dust/water resistance |
Meets IP52, IEC 529 |
Operating altitude |
3000 m above mean sea level (9842 ft) |
Safety Specifications |
Agency approvals |
CAN/CSA C22.2 No 1010.1-92, ASNI/ISA S82.01-1994, UL3111, and EN610-1:1993 |
Mechanical and General Specifications |
Size |
136 x 245 x 63 mm (5.4 x 9.6 x 2.5 in) |
Weight |
1.2 kg (2.7 lb) |
Batteries |
Internal Battery Pack Li-ion: 7.2V,4400mAh, 30 Wh |
Battery life |
>8 hours typical |
Battery replacement |
Replace without opening calibrator; no tools required |
Side port connections |
Pressure module connector |
USB Connector to interface to your PC |
Digital instrument (HART) connector |
Connection for optional battery charger/eliminator |
Data storage capacity |
1 week of calibration procedures results |
90 day specifications |
The standard specification interval for the 750 Seriesare 1 and 2 years. |
Typical 90 day measurement and source accuracy can be estimated by dividing the one year "% of reading" or "% of output" specifications by 2. |
Floor specifications, expressed as "% of full scale" or "counts" or "ohms" remain constant. |